Double row LCD display, front and rear two display screen. Durable stainless steel scale.
Rechargeable battery, rechargeable battery.
Weighing, peeling, counting, unit conversionRS232 data interface, linking computers and printers.
Calibration of 0.01g Precision Balance
(1) Calibration Requirements
When distinct errors appeared in the weighing of balance, the balance shall be recalibrated to make weighing accurate. The recalibrated balance shall be placed on the stable working platform without the influences from the vibration of air current and strong electromagnetic wave. The calibration results will be more accurate 20 minutes later after the balance is operated.
(2) Calibration Procedure
One-point Calibration
Turn on the balance it shall display the weighing mode as “0.00”, then press the “Cal” button until shows “CAL”, then release. Display blinkingly the stand weight value and then place same value standard weight; The “= = = = = =” will be displayed; then display stably weight value and remove weight. then “= = = = = =” will be displayed. the “0.00” will be displayed;. The calibration is finished.
Three-point Calibration
Turn off the balance, then press the “Cal” button and turn on the balance simultaneously until the display shows “CAL”, then release. The calibration procedure shall be undertaken as blow:
Display blinkingly the first point standard weight value then place the same value weight;
Display stably weight value and then remove weight. The “= = = = = =” will be displayed;
Display blinkingly the second point weight value and then place the same value weight;
Display stably weight value and then remove weight. The “= = = = = =” will be displayed;
Display blinkingly the third point stand weight value and then place the same value weight;
Display stably weight value and then remove weight. The “0.00” will be displayed;
the calibration is finished.